Test Code C153 CA 15-3
Specimen and Container/Tube
Specimen Type: Plasma
Preferred Specimen: Li-Heparin Mint Green Top
Other Acceptable specimens: Serum/Gold Top, Serum/Red Top
Specimen Handling and Transport Instructions
UCMC Onsite Instructions:
- Collect specimen per standard collection procedure and send to laboratory immediately.
UCMC Offsite and UC MedLab Outreach Instructions:
- Samples should be centrifuged within 2 hours of collection.
- Centrifuge specimens at >2500 x g for 10 minutes at room temperature.
- Aliquot sample into plastic transport tube containing a minimum of 0.2 mL of specimen.
- Samples must be sent refrigerated at 2-8oC.
Rejection Criteria
- Common specimen rejected situations include: incorrect specimen type, insufficient volume, missing or incomplete specimen identifiers, incorrect specimen transport, or specimens outside stability limits.
- Specimens that are clotted or grossly hemolyzed.
- Specimens that were transported unspun/whole blood and draw time is greater than 6 hours
Reference Values
Test |
Age |
Sex |
Reference Range |
CA 15-3 |
All |
All |
<30 U/mL |
Critical Values
Analyte |
Unit |
Low |
High |
Age |
N/A |
N/A |
N/A |
N/A |
N/A |
CPT Codes
86300
LOINC Codes
6875-9
Volume
Preferred Volume: 0.5 mL
Minimum Volume: 0.2 mL
Specimen Stability
Storage Temperature |
Stability |
Room Temperature |
48 hours |
Refrigerated |
5 days |
Frozen |
90 days |
Collection Instructions
Collect specimen per standard laboratory collection procedures.
Test Components
Cancer Antigen 15-3
Clinical Indications
Immunological in vitro assay for quantitative determination of CA 15-3 in human serum and plasma to aid in the management of breast cancer patients. In conjunction with other clinical and diagnostic procedures, serial testing with the Elecsys CA 15-3 II assay is an aid:
• in the early detection of recurrence in previously treated stage II and III breast cancer patients.
• for monitoring response to therapy in metastatic breast cancer patients.
Methodology
Electrochemiluminescence immunoassay (ECLIA) (Roche Diagnostics)
Additional Information
- Results should always be assessed in conjunction with patient's medical history, clinical examination and other findings.
- The measured tumor marker value of a patient's sample can vary depending on the testing procedure and method used. Tumor marker values determined on patient samples by different testing procedures or methodologies cannot be directly compared with one another and could be the cause of erroneous medical interpretations.
Turnaround Time
Turnaround times are relative to the time the specimen is received in the test laboratory.
STAT Turnaround Time: Not available as STAT test
Routine Turnaround Time: up to 4 hours
Testing Schedule: 24/7
Synonyms/Keyword Search
Cancer Antigen 15-3, Mucin-Like Carcinoma-Associated Antigen
Last Review Date
8/19/2024