Test Code SFPR CSF Protein
Specimen and Container/Tube
Specimen Type: Cerebrospinal fluid
Preferred Specimen: CSF tube
Other Acceptable specimens: Sterile Plastic Container, Sterile Clean Glass Container
Specimen Stability
Storage Temperature |
Stability |
Room temperature |
1 day |
Refrigerated |
6 days |
Frozen |
1 year |
Collection Instructions
- Collect specimen per standard laboratory collection procedures.
- No preservatives or additives.
Specimen Handling and Transport Instructions
UCMC Onsite Instructions:
- Cerebrospinal fluid specimen obtained by clinical procedures.
- Cerebrospinal fluid must be sent to Laboratory Service Center immediately for analysis
- Transport refrigerated (2 - 8ºC).
UCMC Offsite and UC MedLab Outreach Instructions:
- Cerebrospinal fluid specimen obtained by clinical procedures.
- Aliquot specimen into transport tube and freeze.
- Transport refrigerated (2 - 8ºC).
Rejection Criteria
- Common specimen rejected situations include: incorrect specimen type, insufficient volume, missing or incomplete specimen identifiers, incorrect specimen transport, or specimens outside stability limits.
- Specimens that are collected with preservatives or additives.
Reference Values
Test |
Age |
Sex |
Reference Range |
CSF Total Protein |
0 – 1 month |
All |
20 – 80 mg/dL |
CSF Total Protein |
> 1 month old |
All |
15 – 45 mg/dL |
Critical Values
Not applicable
Turnaround Time
Turnaround times are relative to the time the specimen is received in the test laboratory.
STAT Turnaround Time: 1 hour
Routine Turnaround Time: 4 hours
Testing Schedule: 24/7
Clinical Indications
CSF protein measurements are used in the diagnosis and treatment of conditions such as meningitis, brain tumors and infections of the central nervous system.
Test Components
Total Protein, CSF
Methodology
Spectrophotometric, Endpoint (Roche Diagnostics)
CPT Codes
84157
LOINC Codes
TBD
Synonyms/Keyword Search
N/A
Additional Information
Lumbar puncture samples for CSF protein should be collected before fluorescein is given or at least 24 hours later. In a four tube collection series, the third tube is generally designated for chemistry analysis. Therapeutic concentrations of Ca-dobesilate, levodopa and phenazopyridine interfere with the assay. Radiopaque media containing organically bound iodine (e.g. Hexabrix) may cause false-high results. Contamination with blood causes erroneously high CSF protein results.
Last Review Date
4/10/2025
Volume
Preferred Volume: 0.5 mL
Minimum Volume: 0.2 mL