Test Code RTBGPP Blood Gas, Arterial ECMO Pump
Specimen and Container/Tube
Specimen Type: Whole blood
Preferred Specimen: Syringe arterial
Other Acceptable Specimens: None
Specimen Handling and Transport Instructions
UCMC Onsite Instructions:
- Collect specimen per standard collection procedure and send to laboratory immediately.
- Testing must be performed within 30 minutes of collection
UCMC Offsite and UC MedLab Outreach Instructions:
- Specimens are not acceptable from offsite locations for blood gas testing
Rejection Criteria
- Common specimen rejected situations include: incorrect specimen type, insufficient volume, missing or incomplete specimen identifiers, incorrect specimen transport, or specimens outside stability limits.
- Specimens with needles attached will be rejected.
- Specimens with air bubbles.
Reference Values
Analyte |
Age |
Sex |
Normal Range |
SO2, arterial ECMO pump |
All |
All |
94 -98% |
Hemoglobin, arterial ECMO pump |
All |
All |
≥6.0 g/dL |
Methemoglobin, arterial ECMO pump |
All |
All |
≤2.0% |
Critical Values
Analyte |
Age |
Low |
High |
Hemoglobin, arterial |
All |
<6.0 g/dL |
NA |
Methemoglobin |
All |
NA |
>10.0% |
CPT Codes
82805
LOINC Codes
TBD
Volume
Preferred Volume: 0.5 mL
Minimum Volume: 0.2 mL
Specimen Stability
Storage Temperature |
Stability |
Room temperature |
30 minutes |
Refrigerated |
not acceptable |
Frozen |
not acceptable |
Collection Instructions
- Collect specimen per standard laboratory collection procedures.
- Remove all air from syringe.
- Remove needle.
- Cap tightly and mix well.
Test Components
pH, PCO2, PO2, HCO3-, Base Excess, SO2, SAO2, Hemoglobin, Oxyhemoglobin, Deoxyhemoglobin, Carboxyhemoglobin, Methemoglobin
Methodology
Potentiometry: pH, pCO2
Amperometry: pO2
Derived: HCO3-, base excess
Optical (CO-oximetry): SO2, hemoglobin, oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, methemoglobin
Additional Information
- For SO2: Measured oxygen saturation of functional hemoglobin according to the following formula. OxyHb/(OxyHb + DeoxyHb).
- For OxyHb: Measured oxygen saturation of total hemoglobin according to the following formula. OxyHb/(OxyHb + DeoxyHb + COHb + MetHb).
Turnaround Time
Turnaround times are relative to the time the specimen is received in the test laboratory.
STAT Turnaround Time: 1 hour
Routine Turnaround Time: 4 hours
Testing Schedule: 24/7
Last Review Date
4/29/2025