Test Code POCACTHR POC Activated Clotting Time High Range
Specimen and Container/Tube
Specimen Type: Fresh Whole Blood
Preferred Specimen: Plain syringe sample from indwelling line
Other Acceptable specimens: Plain syringe sample from a venipuncture
Specimen Stability
|
Storage Temperature |
Stability |
|
Room temperature |
NA
|
|
Refrigerated |
NA |
|
Frozen |
NA |
Collection Instructions
- Collect at least 0.2 mL of fresh whole blood; avoid drawing from heparinized lines or locks.
- Mix samples thoroughly before applying to the test cartridge.
- Proceed to testing without delay.
Specimen Handling and Transport Instructions
UCMC Onsite Instructions:
Collect whole blood specimen and perform testing immediately.
UCMC Offsite and UC MedLab Outreach Instructions:
NA
Rejection Criteria
Unacceptable Sample Types:
1. Sample contamination with:
- Tissue thromboplastin
- Indwelling intravenous (I.V.) solutions
- Alcohol cleansing solution
- Visible clotting or debris accumulation.
2. Foaming of the sample (air bubbles).
3. Hemolysis.
4. Samples collected in glass blood collection tubes.
5. Unsuspected anticoagulation with either heparin or warfarin.
9. Presence of a lupus anticoagulant.
10.Delay in testing could result in falsely low results.
11. Poor technique during blood collection and delivery to the sample well
Reference Values
|
Test |
Age |
Sex |
Reference Range |
|
ACT Plus |
> 18 years |
All |
82 - 134 sec |
Critical Values
Critical Values:
All ACT tests are performed under the direct supervision of a physician and results are reported immediately.
Turnaround Time
STAT Turnaround Time: N/A
Routine Turnaround Time: N/A
Testing Schedule: 24/7
Clinical Indications
ACT-Plus (High Range Activated Clotting Time) monitors anticoagulation with moderate to high unfractionated heparin (UFH) doses, suitable for use during cardiovascular surgery and cardiac ablation procedures.
Methodology
Clot Detection
CPT Codes
85347
LOINC Codes
96683-8
Synonyms/Keyword Search
Activated Clotting Time High range
ACT High Range
ACT Plus
Last Review Date
06/9/26
Test Components
Activated Clotting Time- High Range
Additional Information
GEM Hemochron 100 ACT test results should be used in conjunction with clinical and laboratory findings.
Test results exhibiting inconsistency with the patient’s clinical status should be repeated or supplemented with additional test data.
Volume
Preferred Volume: 1 mL
Minimum Volume: 200 microliters