Test Code POCACTHR POC Activated Clotting Time High Range

Specimen and Container/Tube

Specimen Type: Fresh Whole Blood

Preferred Specimen: Plain syringe sample from indwelling line

Other Acceptable specimens: Plain syringe sample from a venipuncture

Specimen Stability

Storage Temperature

Stability

Room temperature

NA

 

Refrigerated

NA

Frozen

NA

Collection Instructions

  1. Collect at least 0.2 mL of fresh whole blood; avoid drawing from heparinized lines or locks.
  2. Mix samples thoroughly before applying to the test cartridge.
  3. Proceed to testing without delay.

Specimen Handling and Transport Instructions

UCMC Onsite Instructions: 

Collect whole blood specimen and perform testing immediately.

UCMC Offsite and UC MedLab Outreach Instructions: 

NA

Rejection Criteria

Unacceptable Sample Types:

 

1. Sample contamination with: 

  • Tissue thromboplastin
  • Indwelling intravenous (I.V.) solutions
  • Alcohol cleansing solution
  • Visible clotting or debris accumulation.

2. Foaming of the sample (air bubbles).

3. Hemolysis.

4. Samples collected in glass blood collection tubes.

5. Unsuspected anticoagulation with either heparin or warfarin.

9. Presence of a lupus anticoagulant.

10.Delay in testing could result in falsely low results.

11. Poor technique during blood collection and delivery to the sample well

Reference Values

 

Test

Age

Sex

Reference Range

ACT Plus

> 18 years

All

82 - 134 sec

Critical Values

Critical Values:

All ACT tests are performed under the direct supervision of a physician and results are reported immediately.

Turnaround Time

STAT Turnaround Time: N/A

Routine Turnaround Time: N/A

Testing Schedule: 24/7

Clinical Indications

ACT-Plus (High Range Activated Clotting Time) monitors anticoagulation with moderate to high unfractionated heparin (UFH) doses, suitable for use during cardiovascular surgery and cardiac ablation procedures.

Methodology

Clot Detection

CPT Codes

85347

LOINC Codes

96683-8

Last Review Date

06/9/26

Test Components

Activated Clotting Time- High Range

Additional Information

GEM Hemochron 100 ACT test results should be used in conjunction with clinical and laboratory findings.

Test results exhibiting inconsistency with the patient’s clinical status should be repeated or supplemented with additional test data.

Volume

Preferred Volume: 1 mL

Minimum Volume: 200 microliters