Sign in →

Test Code AHPNL Acute Hepatitis Panel

Important Note

This test may be affected by biotin interference. Patients on high dose biotin (5mg/day or more) should not have samples drawn until at least 24 hours after the last biotin administration.

Infectious

Specimen and Container/Tube

Specimen Type: Serum

Preferred Specimen: Red Top

Specimen Handling and Transport Instructions

UCMC Onsite Instructions: 

Collect specimen per standard collection procedure and send to laboratory immediately. 

 

UCMC Offsite and UC MedLab Outreach Instructions: 

  • Samples should be centrifuged within 2 hours of collection.
  • Centrifuge specimens at >2500 x g for 10 minutes at room temperature.
  • Aliquot sample into plastic transport tube containing a minimum of 0.2 mL of specimen.
  • Samples must be sent room temperature at 15-25oC.

Rejection Criteria

  • Common specimen rejected situations include: incorrect specimen type, insufficient volume, missing or incomplete specimen identifiers, incorrect specimen transport, or specimens outside stability limits. 
  • Specimens that are clotted or grossly hemolyzed.
  • Specimens that were transported unspun/whole blood and draw time is greater than 6 hours

Reference Values


See reference values for Hepatitis A IgM Antibody, Hepatitis B Core IgM Antibody, Hepatitis B Surface Antigen, Heptatis C Antibody

Critical Values

Analyte

Unit

Low

High

Age

N/A

N/A

N/A

N/A

N/A

Volume

Preferred Volume: 0.5 mL

Minimum Volume: 0.2 mL

Specimen Stability

Storage Temperature

Stability

Room Temperature

3 days

Refrigerated

14 days

Frozen

3 months

 

Note: May be frozen up to 6 times.

Collection Instructions

Collect specimen per standard laboratory collection procedures. 

Test Components

Hepatitis A IgM Antibody

Hepatitis B Core IgM Antibody

Hepatitis B Surface Antigen

Heptatis C Antibody

Clinical Indications

Hepatitis A:

Hepatitis A virus (HAV) is an RNA virus that accounts for 20% to 25% of acute viral hepatitis in adults in the United States. Hepatitis A is spread by the oral/fecal route and produces acute hepatitis, which follows a benign, self-limited course. Spread of the disease is usually associated with contaminated food or water caused by poor sanitary conditions. Outbreaks frequently occur in overcrowded situations and institutions or high-density centers such as prisons and healthcare centers. Epidemics may occur following floods or other disaster situations. Chronic carriers of HAV have never been observed.

 

Hepatitis B:

Hepatitis B virus (HBV) is an endemic DNA virus throughout the world. The infection is spread primarily through percutaneous contact with infected blood products (eg, blood transfusion, sharing of needles among injection drug users). The virus is also found in virtually every human body fluid and is known to be spread through oral and genital contact. HBV can be transmitted from mother to child during delivery through contact with blood and vaginal secretions; it is not commonly transmitted transplacentally. After a course of acute illness, HBV persists in approximately 10% of patients. Some chronic carriers are asymptomatic; others develop chronic liver disease, including cirrhosis and hepatocellular carcinoma.

 

Hepatitis C:

Hepatitis C virus (HCV) is an RNA virus that is a significant cause of morbidity and mortality worldwide. The infection is transmitted through contaminated blood or blood products or other close, personal contacts. It is recognized as the cause of most cases of posttransfusion hepatitis. Hepatitis C shows a high rate of progression (~75%) to chronic infection and disease and accounts for the majority of chronic viral hepatitis In the United States. Cirrhosis and hepatocellular carcinoma are sequelae of chronic infection with this virus.

Methodology

Electrochemiluminescence immunoassay (ECLIA) (Roche Diagnostics)

Additional Information

Hepatitis B Surface Antigen Screen:

The Elecsys HBsAg (hepatitis B surface antigen) II assay is based on the sandwich immunoassay principle and performed using an electrochemiluminescence immunoassay on the automated cobas e 801 immunochemistry analyzer. HBsAg present in the patient’s sample reacts with two biotinylated monoclonal anti-HBs, and a mixture of monoclonal anti-HBs and polyclonal anti-HBs labeled with a ruthenium complex react to form a sandwich complex. After addition of streptavidin-coated microparticles (solid phase), the complexes bind to the solid phase via interaction of biotin and streptavidin. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then washed away, and voltage is applied to the electrode that induces chemiluminescent emissions, which are measured by a photomultiplier. Test result is determined by comparing the electrochemiluminescence signal generated from the reaction product in the patient’s sample to the cutoff index (COI) value set from reagent lot-specific assay calibration.(Package insert: Elecsys HBsAG II. Roche Diagnostics; v3.0, 02/2022)

 

Hepatitis B Surface Antigen Confirmation:

The Elecsys HBsAg II Auto Confirm assay is based on the sandwich immunoassay principle and performed using an electrochemiluminescence immunoassay on the automated cobas e 801 immunochemistry analyzer. This test is based on 2 parallel measurements. For the first measurement, the sample is treated with the control pretreatment reagent (PT2) prior to immunoreaction. This measurement serves as a reference. For the second measurement the sample is treated with the confirmatory pretreatment reagent (PT1) prior to immunoreaction. During incubation with confirmatory pretreatment, unlabeled polyclonal anti-HBs are bound to the sample HBsAg and thereby block the binding sites for the labeled antibodies used in the following immunoreaction. The confirmation result (%) is automatically assessed by determining the ratio of both measurements.

 

During testing, the auto-diluted sample is incubated with control pretreatment and confirmatory pretreatment, followed by formation of sandwich complexes of biotinylated monoclonal anti-HBs and a mixture of monoclonal anti-HBs and polyclonal anti-HBs labeled with a ruthenium complex. After addition of streptavidin-coated microparticles (solid phase), the complexes bind to the solid phase via interaction of biotin and streptavidin. The reaction mixture is then aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then washed away, and voltage is applied to the electrode that induces chemiluminescent emissions, which are measured by a photomultiplier. Results are determined by comparing the electrochemiluminescence signal generated from the reaction product to the cutoff index value set from reagent lot-specific assay calibration. The confirmation result (%) is calculated from the ratio of the COI obtained for the measurement with confirmatory pretreatment to the COI obtained for the measurement with control pretreatment.(Package Insert: Elecsys HBsAg II Auto Confirm. Roche Diagnostics; v1.0, 12/2020)

 

Hepatitis A IgM Antibody:

The Elecsys Anti-HAV (hepatitis A virus) IgM assay is based on the sandwich immunoassay principle and performed using an electrochemiluminescence immunoassay on the automated cobas e 801 immunochemistry analyzer. HAV-specific IgM antibody (anti-HAV IgM) in the patient’s serum sample is pretreated with anti-Fdy reagent to block specific IgG in the presence of monoclonal anti-HAV antibodies labeled with ruthenium complex. After addition of biotinylated monoclonal human-IgM-specific antibodies, HAV antigen, and streptavidin-coated microparticles, patient’s anti-HAV IgM form a sandwich complex with the HAV antigen and the ruthenium-labeled anti-HAV antibody which becomes bound to the solid phase via interaction of biotin and streptavidin. The reaction mixture is then aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode, and unbound substances are washed away. Voltage is applied to the electrode and induces chemiluminescent emissions that are measured by a photomultiplier. Test result is determined automatically by the assay-specific software program by comparing the electrochemiluminescence signal generated from the patient’s sample to the COI value set from reagent lot-specific assay calibration.(Package insert: Elecsys Anti-HAV IgM. Roche Diagnostics; v5.0, 11/2022)

 

Hepatitis B Core IgM Antibody:

The Elecsys Anti-HBc (hepatitis B core) IgM assay is based on the sandwich immunoassay principle and performed with an electrochemiluminescence immunoassay on the automated cobas e 801 immunochemistry analyzer. Anti-HBc IgM present in patient’s sample is pretreated with anti-Fdy reagent to block specific IgG. After addition of biotinylated monoclonal human-IgM-specific antibodies, the complexes formed from reaction of ruthenium-labeled HBc antigen, streptavidin-coated microparticles (solid phase), anti-HBc IgM present in the sample, and the biotinylated anti-human IgM bind to the solid phase via interaction of biotin and streptavidin. The reaction mixture is then aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode, and unbound substances are washed away. Voltage is applied to the electrode that induces chemiluminescent emissions, which are measured by a photomultiplier. Test result is determined by comparing the electrochemiluminescence signal generated from the sample to the COI value set from reagent lot-specific assay calibration.(Package insert: Elecsys Anti-HBc IgM. Roche Diagnostics; v1.0, 09/2020)

 

Hepatitis C Virus Antibody:

The Elecsys Anti-HCV (hepatitis C virus) II assay is based on the sandwich immunoassay principle and performed using an electrochemiluminescence immunoassay on the fully automated cobas e 801 immunochemistry analyzer. HCV antibodies present in the patient’s sample react with biotinylated HCV-specific antigens and a reagent containing HCV-specific antigens labeled with a ruthenium complex to form a sandwich complex. After addition of streptavidin-coated microparticles (solid phase), these complexes bind to the solid phase via interaction of biotin and streptavidin. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then washed away, and voltage is applied to the electrode that induces chemiluminescent emissions, which are measured by a photomultiplier. Test result is determined by comparing the electrochemiluminescence signal generated from the patient’s sample to the COI value set from reagent lot-specific assay calibration.(Package insert: Elecsys Anti-HCV II. Roche Diagnostics v1.0, 03/2023)

Turnaround Time

Turnaround times are relative to the time the specimen is received in the test laboratory. 

 

STAT Turnaround Time: Not available as STAT test

Routine Turnaround Time: up to 4 hours

Testing Schedule:  24/7

Last Review Date

06/19/2024